Open original FDA record ↗Recall record
- FDA identifier
- Z-2794-2026
- Company in FDA record
- Karl Storz Endoscopy
- FDA classification date
- Jul 23, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Flexible hysteroscopes indicated for use during diagnostic and operative hysteroscopic procedures may contain an expired single use endoscopic seal for gynecology use, which could impact the assurance of sterility.
Source facts
- Action
- On 5/21/2026, correction notices were emailed to customers who were asked to do the following: 1) Check the expiration date for any inventory of 26153EAU-01 Sealing Seal for Gynecology and discard any expired components(seals) by removing the seal from the packaging and discarding the seal and packaging. 2) Pass on this Urgent Medical Device Correction to all users of affected products and all other persons who need to be aware within your organization. 3) If you have or may have distributed the products listed, please identify and promptly notify those recipients with a copy of this notice and response form. 4) Completed and return the Customer Reply Form via email to KSEAProductLabeling@karlstorz.com For all other questions regarding this communication, please send an email to KSEAProductLabeling@karlstorz.com.
- Initiated
- 2026-05-21
- Root cause
- Packaging change control
- Distribution
- US Nationwide distribution in the states of FL, MA, MN, TX, OH, NY, CA, MI, LA, VA, MD, PA, GA, WI, VT.
- Recall number
- Z-2794-2026
- Recall status
- Open, Classified
- CDRH record ID
- 220593
- FDA classified
- 2026-07-23
- Recall event number
- 99027