Skip to content

Public FDA record

Z-2795-2018: Hamilton-C1 Ventilator Unit Product Usage: The HAMILTON-C1 ventilator is intended to provide positive pr

Recall record Z-2795-2018 for Hamilton Medical AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2795-2018
Company in FDA record
Hamilton Medical AG
FDA record creation date
Aug 14, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An out of specification component on the control board may cause the ventilator to open the inspiratory channel and the expiratory valve. The ventilator will shut down and an emergency buzzer providing an acoustic alarm and the blinking alarm LED on the device s front side.

Source facts

Action
Hamilton Medical notified customers on about 03/09/2018 via letter. Instructions included that a Hamilton Medical Field Service Representative will be in contact with each customer to schedule the correction. Questions or concerns can be directed to Hamilton Technical Support at 1-800-426-6331, option #2.
Initiated
2018-03-09
Root cause
Nonconforming Material/Component
Terminated
2019-03-18
Distribution
US distribution to CA, CO, IL, LA, MN, NY, OK, VA, WI. Worldwide distribution to China, Germany, India, Indonesia, Israel, Kazakhstan, Korea, Pakistan, Taiwan, Poland, Armenia.
Recall number
Z-2795-2018
Recall status
Terminated
CDRH record ID
165558
FDA record created
2018-08-14
Recall event number
80467

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗