Skip to content

Public FDA record

Z-2795-2026: Permobil M3 Corpus Power Wheelchair with Heavy Duty Shock REF:120031, Model version 3

Recall record Z-2795-2026 for Permobil Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2795-2026
Company in FDA record
Permobil Inc
FDA classification date
Jul 24, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

This recall has been reported retroactively. The power wheelchair firmware lacks a speed restriction to maintain stability when the seat lift exceeds 50 mm, which may cause the user to fall from the wheelchair or the wheelchair to tip over,

Source facts

Action
On March 5, 2024, Permobil issued a recall notification to affected consignees via Email. Permobil ask consignees to take the following actions: 1. Do not use the wheelchair when the firmware update is in process. The update is available now and will be proactively sent to the impacted wheelchairs within 1 week. 2. The provider will maintain possession of teh power wheelchair to ensure that the update has successfully installed prior to leaving it with the end user. 3. If not possible, the user should be notified about the pending reported update and instructed to allow this update to complete. 4.If you have any questions or concerns, please contact the Permobil Customer Experience team at (800) 736-0925.
Initiated
2024-02-29
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of CAN.
Recall number
Z-2795-2026
Recall status
Open, Classified
CDRH record ID
221268
FDA classified
2026-07-24
Recall event number
99315

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗