The 00-0903-2510, 3.5mm x 10mm Non Locking Cortical Screw, Lot # 150192-J package may contain a different screw. The screw in the package is a 00-0903-2526, 3.5mm x 26mm, and is etched correctly.
Source facts
Action
OrthoPediatrics has initiated a Voluntary RECALL of Part # 00-0903-2510, 3.5mm Non Locking Cortical Screw, Lot # 150192-J, initially by contacting customers via telephone, and then by e-mailing notification letters on 7/28/2015. Customers returned the 5 mislabled screws, which were repacked with correct labeling. All recall activities were complete as of 8/14/2015.
Initiated
2015-07-28
Root cause
Packaging process control
Terminated
2016-03-17
Distribution
Domestic: TN, IA
Foreign: England
VA/DOD: None
Recall number
Z-2796-2015
Recall status
Terminated
CDRH record ID
139558
FDA classified
2015-09-18
Recall event number
71933