- Recall classification
- Class II
Reason stated in the source
An out of specification component on the control board may cause the ventilator to open the inspiratory channel and the expiratory valve. The ventilator will shut down and an emergency buzzer providing an acoustic alarm and the blinking alarm LED on the device s front side.
Source facts
- Action
- Hamilton Medical notified customers on about 03/09/2018 via letter. Instructions included that a Hamilton Medical Field Service Representative will be in contact with each customer to schedule the correction. Questions or concerns can be directed to Hamilton Technical Support at 1-800-426-6331, option #2.
- Initiated
- 2018-03-09
- Root cause
- Nonconforming Material/Component
- Terminated
- 2019-03-18
- Distribution
- US distribution to CA, CO, IL, LA, MN, NY, OK, VA, WI. Worldwide distribution to China, Germany, India, Indonesia, Israel, Kazakhstan, Korea, Pakistan, Taiwan, Poland, Armenia.
- Recall number
- Z-2796-2018
- Recall status
- Terminated
- CDRH record ID
- 165560
- FDA record created
- 2018-08-14
- Recall event number
- 80467