Open original FDA record ↗Recall record
- FDA identifier
- Z-2798-2026
- Company in FDA record
- AVID Medical, Inc.
- FDA classification date
- Jul 31, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Affected kits contain Medline Namic Angiographic Star Off Handle Manifolds, which were recalled by the manufacturer due to presence of particulates within fluid path of the Manifolds. Particulate within the fluid path has the potential to be introduced into blood circulation and become lodged within blood vessels.
Source facts
- Action
- On April 30, 2026, firm notified distributors via email. Distributors were tasked with notifying their end-users. Customers were instructed to identify, segregate, and quarantine all impacted products. They were provided with a labeling template to print warning labels which should be affixed on all affected kits and cases in inventory. The firm is also able to provide pre-printed labels if needed. Warning labels inform users to remove and discard the manifold prior to use.
- Initiated
- 2026-04-30
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide distribution in the states of NC and TX.
- Recall number
- Z-2798-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221354
- FDA classified
- 2026-07-31
- Recall event number
- 99339