Open original FDA record ↗Recall record
- FDA identifier
- Z-2800-2011
- Company in FDA record
- Cincinnati Sub-Zero Products Inc
- FDA classification date
- Jul 14, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
A component defect. The recalling firm determined that one of the connectors attached to the firm's WarmAir 134 device's Printed Circuit Board (PCB) can overheat. Should this event occur, the excess heat build up within the device may result in damage to the device's external case and/or melting of some of the internal components of the unit.
Source facts
- Action
- CSZ Cincinnati Sub-Zero sent an "URGENT MEDICAL DEVICE RECALL FIELD ACTION" notice to all affected customers. The letter identifies the product, problem, and actions to be taken. CSZ instructs customers to complete and return an enclosed recall response form and to state whether or not their facility is in possession of the device. The completed form should be returned via fax at 513-772-9119 or via e-mail to WA134Recall2011@cszinc.com. Additionally, customers are instructed to discontinue use of the device and return to their Biomedical Engineering and/or Service Department for device exchange and return. Questions regarding this notice are directed to the WarmAir 134 field action (recall) administrator at 1-800-989-7373.
- Initiated
- 2011-05-19
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-02-26
- Distribution
- Worldwide Distribution--USA (nationwide) including the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, LA, MA, MI, MO, NC, NJ, NM, NV, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, and WI and the country of Canada.
- Recall number
- Z-2800-2011
- Recall status
- Terminated
- CDRH record ID
- 100840
- FDA classified
- 2011-07-14
- Recall event number
- 59018