Skip to content

Public FDA record

Z-2800-2020: uMI 550 PET/CT Diagnostic Imaging System - Product Usage: This system is intended to be operated by quali

Recall record Z-2800-2020 for Shanghai United Imaging Healthcare Co., Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2800-2020
Company in FDA record
Shanghai United Imaging Healthcare Co., Ltd.
FDA classification date
Aug 13, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

1) Service function problem may cause false marking of a bad channel resulting in ring artifacts; 2) Potential intermittent scout scanning interruption due to angle signal drift caused by external interference; 3) Metal edge overlap of mylar strip in gantry may create sharp edge. May lead to repeat scanning and abrasions (N/A for Serial No. 200036)

Source facts

Action
Urgent Medical Device Correction Letters regarding the uEXPLORER and uMI 550 were sent to customers on July 16, 2020 via email. Customers were informed that the firm had received complaints on these products and advised customers to take the following actions if problems were observed: 1. if image artifacts occur, reboot the system and continue the reconstruction. If necessary, repeat the scan which will result in an additional dose of radiation and the possible need for additional contrast medium. 2. If scout scanning is unexpectedly interrupted, try repeating the scout. This will result in an additional dose of radiation. 3: Be careful to avoid the sharp edge when the patient raises their arms during movement of the Ml patient table. Customers with questions may contact their United Imaging Service Engineer or call the United Imaging Customer Contact Center at 24-hour hotline (855) 221-1552 (toll free) or (832) 699-2799 (direct). Customers were advised to report adverse reactions or quality problems experienced with the use of these products to the FDA's MedWatch Adverse Event Reporting program either online, by regulatory mail or by fax. Customers were requested to in-turn notify all users of the affected products via the Customer Notification Letter. Customers were informed that United Imaging is updating software to solve these issues, and that United Imaging will replace the mylar strip with an improved one on the affected products via on-site update. Customers are requested to sign the Receipt of Customer Notification Form indicating they received the notice and that they will extend the notification to the final customer. Signed forms are to be returned to the recalling firm.
Initiated
2020-07-16
Root cause
Device Design
Terminated
2022-07-07
Distribution
US Nationwide distribution including in the state of Texas.
Recall number
Z-2800-2020
Recall status
Terminated
CDRH record ID
182610
FDA classified
2020-08-13
Recall event number
85948

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗