Z-2802-2016: RAPIDLab 1260 Blood Gas Analyzer Siemens Material Number (SMN): 10321846, 10491394

Recall record Z-2802-2016 for Siemens Healthcare Diagnostics Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2802-2016
Company in FDA record
Siemens Healthcare Diagnostics Inc
FDA record creation date
Sep 14, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for the first and/or last name of one patient to be printed with Patient ID and result data from a different patient, even though those fields have been turned Off for the system

Source facts

Action
Siemens Healthcare issued Urgent Field Safety Notice to (US) customers via Federal Express. . All affected customers outside the US provided a copy of an Urgent Field Safety Notice via hard copy, e-mail, and/or fax as determined by each countries local regulations and procedures. This notice informs customers of the issue, Actions to be Taken by the Customer and to return the response form. Questions contact your Siemens Customer Care Center or your local Siemens technical support representative
Initiated
2016-08-11
Root cause
Software design
Terminated
2020-07-16
Distribution
Nationwide Foreign: Albania Algeria Angola Argentina Armenia Australia Austria Bahrain Bangladesh Belarus Belgium Bosnia Herzeg. Botswana Brazil Brunei Bulgaria Canada Chile Colombia Croatia Czech Republic Denmark Ecuador Egypt Estonia Finland France Fren.Polynesia Georgia Germany Greece Hong Kong Hungary Iceland India Indonesia Iraq Ireland Israel Italy Ivory Coast Japan Kazakhstan Kenya Kuwait Latvia Lebanon Lesotho Lithuania Luxembourg Macedonia Malaysia Mexico Netherlands New Caledonia New Zealand Norway P.R. China Peru Philippines Poland Portugal Qatar Rep. of Yemen Republic Korea Romania Russian Fed. Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan Thailand Tunisia Turkey Turkmenistan U.A.E. United Kingdom Uruguay Vatikancity Vietnam
Recall number
Z-2802-2016
Recall status
Terminated
CDRH record ID
148764
FDA record created
2016-09-14
Recall event number
74910

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗