Skip to content

Public FDA record

Z-2803-2017: InflammaDry(R) AN RPS DIAGNOSTIC SOLUTION, Sample Collectors InflammaDry(R) is a rapid, immunoassay test

Recall record Z-2803-2017 for Quidel Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2803-2017
Company in FDA record
Quidel Corporation
FDA record creation date
Aug 1, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

In-house testing revealed the presence of plastic spurs at the tips of some Sample Collectors.

Source facts

Action
Quidel sent an Urgent Medical Device Recall letter dated June 22, 2017, to all affected customers. The notice instructed consignees to check inventory for affected lots (11705359 and 1705405), to remove and destroy any affected sample collectors from affected kits, and complete and return the attached Inventory Assessment/Certificate of Destruction form to Quidel by fax at 858-552-7905 or email to technicalsupport@quidel.com. Any questions by consignees should be directed to Quidel's Technical Support Team at 800-874-1517 or 858-552-1100, or by email at technicalsupport@quidel.com. It is requested that consignees keep recall notification and response in their records.
Initiated
2017-06-22
Root cause
Process design
Terminated
2019-06-07
Distribution
Nationwide Distribution including AR, AZ, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, NC, NE, NH, NJ, NY, OH, OK, PA, SC, SD, TX, VA, WA, & WV.
Recall number
Z-2803-2017
Recall status
Terminated
CDRH record ID
157032
FDA record created
2017-08-01
Recall event number
77640

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗