Open original FDA record ↗Recall record
- FDA identifier
- Z-2803-2017
- Company in FDA record
- Quidel Corporation
- FDA record creation date
- Aug 1, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
In-house testing revealed the presence of plastic spurs at the tips of some Sample Collectors.
Source facts
- Action
- Quidel sent an Urgent Medical Device Recall letter dated June 22, 2017, to all affected customers. The notice instructed consignees to check inventory for affected lots (11705359 and 1705405), to remove and destroy any affected sample collectors from affected kits, and complete and return the attached Inventory Assessment/Certificate of Destruction form to Quidel by fax at 858-552-7905 or email to technicalsupport@quidel.com. Any questions by consignees should be directed to Quidel's Technical Support Team at 800-874-1517 or 858-552-1100, or by email at technicalsupport@quidel.com. It is requested that consignees keep recall notification and response in their records.
- Initiated
- 2017-06-22
- Root cause
- Process design
- Terminated
- 2019-06-07
- Distribution
- Nationwide Distribution including AR, AZ, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, NC, NE, NH, NJ, NY, OH, OK, PA, SC, SD, TX, VA, WA, & WV.
- Recall number
- Z-2803-2017
- Recall status
- Terminated
- CDRH record ID
- 157032
- FDA record created
- 2017-08-01
- Recall event number
- 77640