Skip to content

Public FDA record

Z-2803-2026: L-Cath PICC S/L Peripherally Inserted Central Catheter: 26ga (1.9F) 0.60mm x 30cm, REF: 384539; 18ga (3.5

Recall record Z-2803-2026 for Argon Medical Devices, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2803-2026
Company in FDA record
Argon Medical Devices, Inc
FDA classification date
Jul 24, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Catheters, indicated for short/long-term peripheral access to central venous system for the administration of fluids, medications, nutrients; and sampling, may experience leaking and/or cracking at the hub, which may lead to replacement, inability to flush/aspirate catheter, catheter occlusion, therapy interruption, blood loss, tissue irritation, bloodstream infection.

Source facts

Action
On 6/15/2026, recall notices were emailed to customers who were asked to do the following: 1) Provide this recall notice to all people who need to be aware within your organization or to any organization where the potentially affected product has been transferred. 2) Segregate, quarantine, and return affected devices to recalling firm. 3) Complete and return the response form via email Arbee.cummings@argonmedical.com For additional information on this letter, please email Scott.Bishop@argonmedical.com
Initiated
2026-06-15
Root cause
Device Design
Distribution
US: AR, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, MA, MD, MI, MN, MO, MS, NC, ND, NJ, NV, NY, OR, TN, TX, UT, VA, WA. OUS: MYS, CAN, SAU
Recall number
Z-2803-2026
Recall status
Open, Classified
CDRH record ID
221315
FDA classified
2026-07-24
Recall event number
99333

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗