Open original FDA record ↗Recall record
- FDA identifier
- Z-2805-2026
- Company in FDA record
- Permobil Inc
- FDA classification date
- Jul 24, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
This recall is being reported retroactively. Power Platform firmware is missing specific drive inhibit logic. This issue may allow the chair to drive under restricted conditions, and may create an unstable condition in certain circumstances.
Source facts
- Action
- On June 10, 2025 Permobil issued a technical bulletin (recall notice) via Email to affected consignees. Permobil informed consignees of the following: 1. Cellular connection is required to receive and install the firmware update. 2. It is important to not use the wheelchair when the firmware update is in process (as indicated on the display of the input device). 3. Specific Remote Update Instructions are available if needed in addition to the on-screen information. 4. A confirmation of the completed update will be shown on the joystick display when the update successfully completes. 5. If you have any questions or concerns, please contact the Permobil Customer Experience team at (800) 736-0925. 1.
- Initiated
- 2025-06-19
- Root cause
- Process design
- Distribution
- Worldwide - US: Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of CANADA.
- Recall number
- Z-2805-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221171
- FDA classified
- 2026-07-24
- Recall event number
- 99267