Z-2807-2015: EVIS EXERA II Duodenovideoscope OLYMPUS TJF TypeQ180V. This instrument has been designed to be used wit

Recall record Z-2807-2015 for Olympus America Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2807-2015
Company in FDA record
Olympus America Inc.
FDA classification date
Sep 22, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

New and validated reprocessing procedures have been issued for the Olympus TJF-Q180V duodenoscope ("TJF-Q180V"), consisting of revised manual cleaning and high level disinfection procedures.

Source facts

Action
The firm, Olympus, sent an "Urgent Safety Notification" dated March 26, 2015, to direct consignees/customers. The notification described the product, problem and actions to be taken. The notification contained new reprocessing procedures. The customers were instructed to immediately implement new precleaning and high level disinfection procedure; ensure that all personnel involved in the reprocessing are completely knowledgeable and thoroughly trained on the new reprocessing instructions; complete and return the enclosed questionnaire via fax to (484) 896-7128 and implement the new cleaning instructions when the new brushes arrive on/about May 8, 2015. If you have question, contact our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, Monday-Friday between 7 AM EST - 8 PM EST. For additional information on this matter, contact V.P., Regulatory/Clinical Affairs & Quality Assurance at (484) 896-5688 or by email at laura.storms@olympus.com.
Initiated
2015-03-26
Root cause
Labeling design
Terminated
2017-02-01
Distribution
Worldwide Distribution: US (nationwide) and countries of: Canada and Mexico.
Recall number
Z-2807-2015
Recall status
Terminated
CDRH record ID
135241
FDA classified
2015-09-22
Recall event number
70858

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗