Skip to content

Public FDA record

Z-2810-2020: Phillips 3.7V Rechargeable Li-Polymer Battery, Product #: 989803191341 - Product Usage: This battery is u

Recall record Z-2810-2020 for Philips North America, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2810-2020
Company in FDA record
Philips North America, LLC
FDA record creation date
Aug 14, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A problem has been detected with the Philips LI-Polymer rechargeable battery when used with its approved charging station and module charger. When the battery is inserted into a charger bay, the LED indicatory lights solid red and the battery will not charge. This may result from poor electrical contact between the battery and charger.

Source facts

Action
Philips Healthcare sent a customer notification letter dated June 2020 explaining the potential issues observed with their products. This letter instructs the consumer to contact the company with any reports of issues associated with this failure.
Initiated
2020-06-18
Root cause
Nonconforming Material/Component
Terminated
2023-01-10
Distribution
World wide distribution. Firm reports distributing more product then quantity manufactured. Clarification requested. And received. Firm made error in counting number of batteries distributed.
Recall number
Z-2810-2020
Recall status
Terminated
CDRH record ID
182528
FDA record created
2020-08-14
Recall event number
86025

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗