Open original FDA record ↗Recall record
- FDA identifier
- Z-2810-2020
- Company in FDA record
- Philips North America, LLC
- FDA record creation date
- Aug 14, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A problem has been detected with the Philips LI-Polymer rechargeable battery when used with its approved charging station and module charger. When the battery is inserted into a charger bay, the LED indicatory lights solid red and the battery will not charge. This may result from poor electrical contact between the battery and charger.
Source facts
- Action
- Philips Healthcare sent a customer notification letter dated June 2020 explaining the potential issues observed with their products. This letter instructs the consumer to contact the company with any reports of issues associated with this failure.
- Initiated
- 2020-06-18
- Root cause
- Nonconforming Material/Component
- Terminated
- 2023-01-10
- Distribution
- World wide distribution. Firm reports distributing more product then quantity manufactured. Clarification requested. And received. Firm made error in counting number of batteries distributed.
- Recall number
- Z-2810-2020
- Recall status
- Terminated
- CDRH record ID
- 182528
- FDA record created
- 2020-08-14
- Recall event number
- 86025