Z-2814-2020: Smart Stapes Protheses Piston 0.6mm X 4.25 mm- Partial Ossicular Replacement Catalog Number: 70143662

Recall record Z-2814-2020 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2814-2020
Company in FDA record
Olympus Corporation of the Americas
FDA record creation date
Aug 14, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Mislabeled outer carton: packaged with the outer carton label stating the contents were the 70143662 Smart Piston but the inside package contents was the 70142158 Stapes Piston; see image below

Source facts

Action
Olympus issued Urgent Medical Device Removal Action letter, dated July 14, 2020 via Federal Express stating reason for recall, health risk and action to take: 1.Immediately assess any affected product you have in stock and quarantine any affected product. 2.Access the Olympus recall portal to indicate that you have received this notification. Go to https://olympusamerica.com/recall. Enter the recall number Recall-XXXX and provide your contact information as indicated in the portal. 3. Olympus will issue a credit or replacement for your affected inventory. In addition, if you may have further distributed this product, please identify your customers, notify them at once of this field action and appropriately document your notification process. Your notification to your customers may be enhanced by including a copy of this field action notification letter. Olympus regrets any inconvenience caused and fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me directly at (484)896-5688 from Monday till Friday 8AM ET to 5PM ET or by e-mail at laura.storms@olympus.com.
Initiated
2020-07-14
Root cause
Packaging process control
Terminated
2021-09-07
Distribution
CA , FL , GA , PA Foreign: New Zealand , Switzerland
Recall number
Z-2814-2020
Recall status
Terminated
CDRH record ID
182562
FDA record created
2020-08-14
Recall event number
86045

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗