Z-2816-2020: DePuy Synthes Trauma LCP One-Third Tubular Plate with Collar 8 Holes / 93mm - Product Usage: intended for

Recall record Z-2816-2020 for Synthes (USA) Products LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2816-2020
Company in FDA record
Synthes (USA) Products LLC
FDA record creation date
Aug 17, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

May have one or more dimensional features out of tolerance includes: thread pitch diameter depth, thread minor diameter depth, plate thickness and end length out of tolerance that can potentially result in a reduced locking strength of an LCP screw

Source facts

Action
DePuy Synthes issued Urgent Medical Device Removal letter dated July 21, 2020 stating reason for recall, health risk and action to take: Please take the following actions: 1. Immediately examine your inventory to determine if you have product subject to this action on hand and quarantine the subject product. 2. Complete the attached Verification Section (page 3 of this letter). Please include your facility name and address, account number, name of person completing the form, title, email address, telephone number and signature in the spaces provided. If you have product to be returned: " Call DePuy Synthes at 1-800-479-6329 to obtain a Return Authorization (RA) Number. " Return the completed Verification Section (page 3 of this letter) with the product to: Credit/Returns, DePuy Synthes, 1101 Synthes Avenue, Monument, CO 80132 " Send a copy of the completed Verification Section (page 3 of this letter) to DePuy Synthes by: Scan/email: RA-DPYCH-flactionint@ITS.JNJ.com " Return any subject product within 10 business days. If you have no product to be returned: " Return the Verification Section (page 3 of this letter) to DePuy Synthes by: Scan/email: RA-DPYCH-flactionint@ITS.JNJ.com 3. Forward this notice to anyone in your facility that needs to be informed. 4. If any of the subject product has been forwarded to another facility, contact that facility to arrange return. 5. Keep a copy of this notice visibly posted for awareness until the product subject to this action has been returned. While processing your returns, please maintain a copy of this notice with the product subject to this action and keep a copy for your records. Questions, please call 610-719-5450 or contact your DePuy Synthes Sales Consultant
Initiated
2020-07-21
Root cause
Process control
Terminated
2022-02-18
Distribution
US Nationwide distribution.
Recall number
Z-2816-2020
Recall status
Terminated
CDRH record ID
182662
FDA record created
2020-08-17
Recall event number
86104

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗