In affected software version, of Positron emission tomography and computed tomography system, the CT Hounsfield unit curve may not display in real time when using the bolus tracking protocol with the tracker scan function. If this occurs, the subsequent clinical scanning protocols will need to be started manually, which may effect image enhancement, and may result in patient rescan.
Source facts
Action
On March 26, 2020, Customer Notification Letters were emailed to customers.
The recalling firm has released a software hotfix to address the issue for the affected products. The corrective action of on-site upgrade will be provided free of charge. For further questions, feel free to contact the Customer Contact Center at (855) 221-1552 (toll free) or (832) 699-2799 (direct).
Initiated
2020-03-26
Root cause
Software design
Terminated
2023-08-02
Distribution
US Nationwide distribution including in the state of TX.
Recall number
Z-2819-2020
Recall status
Terminated
CDRH record ID
180937
FDA record created
2020-08-18
Recall event number
85453