Skip to content

Public FDA record

Z-2820-2020: LDI Laser firmware lower than V02.13.00 for LDI models: LDl-7, LDI-WF, LDI-NIR, LDl-6 Laser Science flu

Recall record Z-2820-2020 for 89 NORTH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2820-2020
Company in FDA record
89 NORTH
FDA record creation date
Aug 25, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The defect relates to firmware version lower than V02.13.00. It is controlling safety interlocks of the LDI; specifically, the potential for the LDI to enter a service mode intended for use in the factory environment only which overrides all interlocks. When this mode - "Interlock Override Mode" - is entered, the system does not cease laser emission when interlocks are opened, but rather allows emission to continue with open interlocks. The potential for this to occur is non-compliant with 21CFR 1040 .10{f )(2) and 21 CFR 1040 .10(f)(3).

Source facts

Action
All customers who purchased an LOI prior to identification of the defect were notified of the defect electronically per the email address on record for each customer. Each affected LOI was identified by serial number per our production records. 2. The firmware upgrade was distributed to all affected customers via the electronic notification. A website link embedded in the electronic notification brings customers to an electronic form where they are asked to submit their contact information as well as the serial numbers of the LDI units they are upgrading. 3. Upon completing the form, the user is provided (via the form) another link which contains the firmware download as well as instructions for performing the upgrade. 4. For those customers who want assistance (as stated in the notification to customers), 89 North technical support can perform the upgrade for them via a remote session to the customer's computer which is interfaced to the LDI product. Please contact us with any questions or if you need any assistance completing the update and we would be happy to help you. Feel free to reach out to your normal 89 North support contact, or you can reach us at (802) 881-0302 (menu option 6) or ldi-fw-supoort@89north.com.
Initiated
2020-01-31
Root cause
Radiation Control for Health and Safety Act
Terminated
2023-05-02
Distribution
Worldwide
Recall number
Z-2820-2020
Recall status
Terminated
CDRH record ID
182866
FDA record created
2020-08-25
Recall event number
86158

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗