Reason stated in the source

Modular Junction Box (MJB) cables can melt and affect the bed functionality.

Source facts

Action
On July 21st 2020 the firm, ARJO, sent an "URGENT-FIELD SAFETY NOTICE" notification to their US customers indicating the following: Although the likelihood of a serious injury is limited, (as the specific sequence of events is necessary), Arjo is taking this matter very seriously and intends to perform corrective action on the affected units. Next Steps 1. Verify whether you possess devices within the range of this Notice and ensure that all caregivers and users of Arjo Citadel Plus beds are made aware of this Field Safety Notice (FSN). If you do not use Skin IQ mattress, check the position of the Skin IQ cable (situated at the footend of the bed). In order to mitigate the risk of electric shock occurrence, protect the cable connector against touching, e.g by hiding it under the mattress. If Skin IQ mattress is being used (is connected via cable) no risk of electrick shock is present. 3. Ensure that the device does not stay in use if the malfunction occurs (control panels starts blinking, the functionality of the bed is disturbed, unpleasant smell is emitted), until the repair is performed. 4. Complete and sign the enclosed Customer Response Form and return this form to your local Arjo office; Fax: 1-833-329-2756, email:mailto:ArjoCustomer.Response-US@Arjo.com or mail: Arjo Inc., 2349 W Lake St. #250, Addison IL 60101. Once returned, Arjo will contact the designated facility representative listed on the Customer Response Form to schedule the service visit and repair all affected devices at your facility. In situations where Arjo is already aware of number of impacted units at your location, Arjo will offer immediate repair or swap to avoid any logistical inconvenience to our customers. Note: if your facility has sold or moved affected Citadel Bed, please include the new facilitys information in the Customer Response Form. The notice has been submitted to the Regulatory Agency/National Competent Authority in your country. If you have any furth
Initiated
2020-07-21
Root cause
Device Design
Terminated
2021-02-17
Distribution
Worldwide distribution: US (nationwide) to states of:: FL, KS, LA and countries of:: Australia, Brazil, Canada, France and Germany.
Recall number
Z-2821-2020
Recall status
Terminated
CDRH record ID
182593
FDA record created
2020-08-18
Recall event number
86065

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗