Open original FDA record ↗Recall record
- FDA identifier
- Z-2826-2015
- Company in FDA record
- Ferno-Washington Inc
- FDA classification date
- Oct 1, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Hole elongation in the telescoping load frame channel. A hole elongation could affect the ability of the retention pin to stay seated which is necessary to preclude movement of the telescoping load frame.
Source facts
- Action
- Ferno began calling their customers on 8/3/2015. The firm informed their customers of the following: To provide optimal Customer Experience and minimal disruption to your service, the iNX field action involves the following steps: " Exchange of your current iNX with a new iNX; " Participation in a brief training session; and " Inspection of your Ferno Model 175 antler and rail fastening system as requested by the Dept. of Transportation. As you await delivery of your new iNX, it is important you ensure proper function of the fastening system and iNX telescoping load frame.
- Initiated
- 2015-08-03
- Root cause
- Device Design
- Terminated
- 2016-01-20
- Distribution
- Worldwide Distribution - US Nationwide in the states of: CO, IA, IL, IN, KY, MD, MN, MO, NC, NJ, NY, OH, OK, OR, TX, VA, WA & WY and the countries of CANADA, AUSTRALIA, UNITED KINGDOM & NEW ZEALAND.
- Recall number
- Z-2826-2015
- Recall status
- Terminated
- CDRH record ID
- 139518
- FDA classified
- 2015-10-01
- Recall event number
- 71983