Z-2828-2020: NobelReplace Conical Connection (CC) PMC RP 4.3x10mm, REF: 37291 - Product Usage: implants are intended t

Recall record Z-2828-2020 for NOBEL BIOCARE SERVICES AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2828-2020
Company in FDA record
NOBEL BIOCARE SERVICES AG
FDA record creation date
Aug 20, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Dental implant cap label may incorrectly indicate implant measurements 5.0x8mm. Actual implant measurements are 4.3x10mm. If not noticed prior to implant placement, this could lead to: 1) A second surgical procedure; 2) Implant replacement; 3) If placed in posterior lower jaw, the implant may damage the inferior alveolar nerve canal or perforate a small artery requiring immediate implant removal.

Source facts

Action
On 07/20/20, Urgent Medical Device Recall notices were emailed to customers who were asked to do the following: 1. Inspect your stock and confirm that both the cap label and the vial label indicate 4.3x10mm. 2. Quarantine devices with cap label 5.0x8mm. 3. Complete attached Customer Acknowledgment Form, even if you do not have any affected stock, and return it to the recalling firm via email. 4. Return all incorrectly labeled devices to the firm. Correctly labeled devices do not need to be returned. 5. Ensure relevant staff members are informed of this recall. If you have supplied or transferred any potentially affected product to another facility or organization, let that facility know of the recall by providing a copy of this letter. 6. Firm will replace returned affected devices. If you require any further information or support, please contact your local customer support representative at 714-282-4800 ext. 1527.
Initiated
2020-07-20
Root cause
Process control
Terminated
2021-05-12
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of FL and NY and the countries of Albania and Hong Kong.
Recall number
Z-2828-2020
Recall status
Terminated
CDRH record ID
182623
FDA record created
2020-08-20
Recall event number
86085

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗