Z-2830-2020: Ultra/Phonic Focus BP Conforming Gel Pads, UDI 00300361301021 (Primary); 10300361301028 (Package) - Produ

Recall record Z-2830-2020 for Pharmaceutical Innovations, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2830-2020
Company in FDA record
Pharmaceutical Innovations, Inc.
FDA record creation date
Aug 20, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm registered the product as a Class I device, but has since been informed by FDA that the device is not Class I and requires a 510(k) submission.

Source facts

Action
On 8/13/2020, the firm distributed Urgent Medical Device Recall letters to its distributor, with instructions to notify all affected customers who received the product. Customers were informed that the product had been misclassified by the firm and required a 510(k) submission to FDA. Customers were instructed to identify and quarantine the products, and to cease use and sales of the product. Customers are to complete the included response form and return it to the firm. The firm will issue a Return Material Authorization for the return of the product. No option will be given to destroy product in the field. Customers should notify all users within their facility as well as any downstream accounts if the product has been further distributed/ transferred.
Initiated
2020-08-13
Root cause
No Marketing Application
Terminated
2021-09-16
Distribution
US Nationwide distribution including in the state of Nevada.
Recall number
Z-2830-2020
Recall status
Terminated
CDRH record ID
182784
FDA record created
2020-08-20
Recall event number
86138

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗