Open original FDA record ↗Recall record
- FDA identifier
- Z-2830-2026
- Company in FDA record
- Stryker GmbH
- FDA classification date
- Jul 29, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization.
Source facts
- Action
- Firm began notifying consignees on July 1, 2026. Customers were instructed to immediately check their internal inventory to locate affected product. Customers were informed that until further notice, the subject devices are limited to single use only. Following first use, discard the devices in accordance with the facility's established procedures. Users should conduct visual inspection of product prior to each use. If cracking is found to be present, the customer should quarantine non-confirming parts and report the issue to Stryker's Trauma & Extremities division. Rods with cracks should be returned as part of the complaint process.
- Initiated
- 2026-07-01
- Root cause
- Device Design
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, ID, IL, KS, LA, MA, MD, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NY, OH, PA, RI, SC, TN, TX, VA, WI, WV and the countries of Australia, Austria, Belgium, Canada, Chile, Denmark, France, Germany, Greece, Ireland, Israel, Italy, Japan, Kuwait, Netherlands, Poland, Portugal, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, UK.
- Recall number
- Z-2830-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221139
- FDA classified
- 2026-07-29
- Recall event number
- 99246