Skip to content

Public FDA record

Z-2833-2016: TriFoillmaging Triumph/Triumph II Research CT scanners.

Recall record Z-2833-2016 for Northridge Tri-Modality Imaging,. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2833-2016
Company in FDA record
Northridge Tri-Modality Imaging,
FDA record creation date
Sep 30, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Firm has determined that several Triumph/Triumph II Research CT scanners produced by Northridge Trimodality Imaging, inc dba TriFoillmaging (or by Gamma Medica-Ideas, Inc and field upgraded with a replacement x-ray tube) may be out of compliance with 21 CFR 1020.40 (Cabinet X-ray standards)

Source facts

Action
TriFoil lmaging planned action to bring defect into Compliance: 1. All repairs were performed at customer facilities. Customers were contacted by phone to arrange a convenient repair time. 2. TriFoil service engineers installed a collimator on each of the affected cabinet x-ray systems to bring their emission rates into compliance with the emission limit in the performance standard. 3. Full radiation surveys were performed, at maximum exposure settings, to verify each repair was effective. 4. All costs associated with the repair and follow-up testing were paid directly by TriFoil Imaging. CDRH approves the CAP and understands that the CAP was completed on August 29, 2016. For further questions please call (818) 709-2468.
Initiated
2016-05-24
Root cause
Radiation Control for Health and Safety Act
Terminated
2016-10-06
Distribution
USA Distribution
Recall number
Z-2833-2016
Recall status
Terminated
CDRH record ID
148565
FDA record created
2016-09-30
Recall event number
74838

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗