Software issue. The difference between the display and dosimeter readings In the Child/Pediatric automatic exposure mode is in the range of 52-65%. The display in all other modes deviated by less than 35% from the dosimeter readings.
Source facts
Action
The firm, Liebel-Flarsheim, notified its customers by phone on or about 2/27/2017 to inform them of the product, problem and actions to be taken. The firm plans to have these Sedecal software updates implemented over the course of the next 12 months.
For additional information, please contact me at 314-376-4788.
Initiated
2017-02-27
Root cause
Radiation Control for Health and Safety Act
Terminated
2023-08-16
Distribution
US Distribution to: AL, AZ, CA, CT, CO, FL GA, HI, IA, IL, ID, KS, KY, MD, MI, Michigan, MN, NC, NH, NV, NY, OK, OH, PA, SC, TN, TX, VA, WA, WI, and WV,
Recall number
Z-2834-2018
Recall status
Terminated
CDRH record ID
164770
FDA record created
2018-08-21
Recall event number
80174