An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may result in the Endoscope Controller showing surgical images that are dimmer, have color imbalance or have reduced fluorescence and grayscale when using Firefly. There is also potential for increased fogging of the endoscope.
Source facts
Action
Intuitive Surgical notified customers via a "Field Safety Notice
Urgent Medical Device Correction ISIFA2018-06-C
da Vinci Xi Miscalibrated Endoscope Controller" letter. The letter provided: Reason for Field Action, Risk to Health, Affected Products (description, part number, serial number), Action taken by Intuitive Surgical, Actions to be taken by the Customer/ User, contact information for Further Information & Support, and an Acknowledgement Form.
Initiated
2018-05-17
Root cause
Use error
Distribution
U.S.
Recall number
Z-2835-2018
Recall status
Open, Classified
CDRH record ID
166162
FDA record created
2018-08-16
Recall event number
80691