Open original FDA record ↗Recall record
- FDA identifier
- Z-2838-2018
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- Aug 17, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
In rare cases of insufficient maintenance or high clinical workload, the first rope of the ceiling stand, which is designed to take the load, could break without triggering the safety lock and this could lead to overloading the safety rope which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down during patient positioning. Patients and users may be seriously injured.
Source facts
- Action
- On 6/29/18, the firm emailed its four affected original equipment manufacturers (OEMs), alerting them to the potential malfunction of the ceiling stand's lifting column that could cause the arm to drop down during patient positioning. Manufacturers were asked to distribute an attached Safety Notice to affected customers. In addition, the firm urged manufacturers to order retrofit kits for all customers and to take any steps needed to ensure that customers install the kits. The kits are free of charge. Manufacturers were also asked to complete a response form and return to the firm. Affected customers were advised of the product problem and instructed to install retrofit kits to avoid risk to patients and to users. If you have any questions, please contact Mr. Wolfgang Bohrer at wolfgang.bohrer@siemens-healthineers.com.
- Initiated
- 2018-06-29
- Root cause
- Device Design
- Terminated
- 2019-10-04
- Distribution
- US Nationwide Distribution to accounts in IL, FL, and NY.
- Recall number
- Z-2838-2018
- Recall status
- Terminated
- CDRH record ID
- 166243
- FDA record created
- 2018-08-17
- Recall event number
- 80735