Open original FDA record ↗Recall record
- FDA identifier
- Z-2839-2018
- Company in FDA record
- Hitachi Medical Systems America Inc
- FDA record creation date
- Aug 17, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm received a complaint where, during patient positioning, the technologist was adjusting the patient table height and the table failed to stop when the control was released and contacted the CT gantry. The problem could not be duplicated at the site, so to avoid further issues while a root cause investigation was pursued, Hitachi choose to add an optional touch sensor to the gantry to add a redundant level of protection. It is conceivable that a patient s extremity may be caught between the tabletop and the gantry surface during uncontrolled motion or that the table contact of the gantry might cause internal damage to the unit.
Source facts
- Action
- On 6/19/18, the firm was notified of the problem at a customer site. The firm is currently investigating the underlying cause to formulate a planned corrective action. On 6/29/18, the firm took a preventive measure at the site which originated the complaint: Hitachi added an optional touch sensor to the unit to lessen the chance of injury. The firm may take additional corrective action on other units distributed to customers once root cause investigation is complete.
- Initiated
- 2018-06-29
- Root cause
- Nonconforming Material/Component
- Terminated
- 2020-04-27
- Distribution
- US Distribution to one customer in TX.
- Recall number
- Z-2839-2018
- Recall status
- Terminated
- CDRH record ID
- 166120
- FDA record created
- 2018-08-17
- Recall event number
- 80671