Skip to content

Public FDA record

Z-2840-2016: G1 Dissection Tool - 6MM CRS DBALL, 2mm Shaft EXP, M; For cutting and shaping bone including spine and cr

Recall record Z-2840-2016 for The Anspach Effort, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2840-2016
Company in FDA record
The Anspach Effort, Inc.
FDA record creation date
Sep 16, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm discovered on March 4, 2016 that several product codes were inadvertently left off of Hold 1503.

Source facts

Action
On April 26, 2016 DePuy Synthes sent a letter informing their customers of the recall. ACTION REQUIRED Our records show that your facility has received one or more of product(s) subject to the recall (correction). The Anspach Effort, Inc. asks that you: 1) There is no requirement for retaining your inventory of Gl Dissection Tools. 2) To ensure Gl Cutting Tool & Attachment compatibility, please reference Table 1 to review which Attachments are compatible. The Cutting TooVAttachment combination noted with a check mark in Table I will function as intended. The Cutting Tool/Attachment combinations noted with an "X" in Table 1 will not secure effectively and should not be used. 3) Review, complete, sign and return the attached reply form confirming your receipt of this letter by fax or email details provided on the form. Returning the form promptly will prevent you from receiving repeat notices. 1. Share this letter with others in your facility that needs to be made aware of this recall (correction). 2. If you distribute any of the product(s) subject to the recall (correction) to other services or facilities, please forward this letter as appropriate immediately. 3. Keep a copy of this notice with the product(s) subject to the recall (correction). If you have any questions regarding this correction, please contact the Complaint Handling Unit Manager at 561-494-3673 or contact your Anspach Sales Representative.
Initiated
2016-08-26
Root cause
Device Design
Terminated
2017-11-29
Distribution
AZ, DC, CA, MD, MI, and OR Switzerland
Recall number
Z-2840-2016
Recall status
Terminated
CDRH record ID
149429
FDA record created
2016-09-16
Recall event number
75121

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗