Z-2840-2018: ELEKTA Digital Accelerator under the following brand names: Elekta Synergy, Synergy S, Synergy Platform,

Recall record Z-2840-2018 for Elekta, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2840-2018
Company in FDA record
Elekta, Inc.
FDA record creation date
Aug 17, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for an uncontrolled extension of iViewGT / XVI detector arms.

Source facts

Action
On December 6, 2017 Elekta sent notification to their customers informing them the following: Description of Problem: Elekta has identified that if the microswitch within the middle arm is not set up correctly, it is possible for the MV or kV detector arm to extend due to gravity. When the detector arm is fully open there is no risk. Details: In order for an uncontrolled extension to occur on a detector arm, the following conditions must be present: The microswitch (within the Solenoid Limit Switch of the PCB assembly in the middle arm) is activated incorrectly. The detector-locked solenoid is not fully engaged in the interface bracket. This is most likely to lead to a potential collision in the following scenarios: iViewGT - The detector arm is opened or closed above the patient, when the gantry angle is at 180 ( 45). XVI - The detector arm is opened or closed above the patient, when the gantry angle is at 90 ( 45) Clinical Impact: It is possible that a collision with the extended detector arm can cause serious injury. Recommended User Action: Elekta recommends that the user does not open or close the detector arm when it is positioned above the patient, as specified in the Instructions For Use (IFU). Elekta Corrective Actions: Elekta Field Service Representatives will do checks to make sure that that the microswitch activates only when the detector-locked solenoid is fully engaged in the interface bracket. If you have any concerns, contact your Elekta Field Service Representatives who are aware of how to do these checks. This notice has been provided to the appropriate Regulatory Authorities. We sincerely apologize for any inconvenience this action may cause and thank you in advance for your cooperation.
Initiated
2017-12-01
Root cause
Component design/selection
Terminated
2019-11-05
Distribution
Worldwide Distribution - US Nationwide in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, Alaska, District of Columbia, Hawaii and Puerto Rico Albania, Algeria Antigua and Barbuda Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Gabon, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran (Islamic Republic of), Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Lao People's Democratic Republic, Libyan Arab Jamahiriya, Lithuania, Malaysia, Malta, Mexico, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Panama, Paraguay, Peru, Philippines, Portugal, Republic of Serbia, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Kingdom, Venezuela, Viet Nam and Zimbabwe
Recall number
Z-2840-2018
Recall status
Terminated
CDRH record ID
166161
FDA record created
2018-08-17
Recall event number
80690

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗