Open original FDA record ↗Recall record
- FDA identifier
- Z-2848-2020
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Aug 22, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Hardware error- Hot Plugging feature could cause an electrical malfunction on the main control D80 (below revision 04 installed) of the generator which is part of the main unit (c-arm) and cause permanent loss of imaging functionality.
Source facts
- Action
- Siemens Medical Solutions USA, Inc. initiated a Customer Safety Advisory Notice to affected customers via AX013/20/S dated July 14, 2020, The letter states reason for recall, health risk and action to take: We advise urgently to cease the use of the hot plugging feature until the pending hardware update AX014/20/S has been completed. Siemens will correct the hardware error via Update Instruction AX014/20/S in which a Siemens Service Engineer will replace the main control D80. Service organization will contact you shortly to arrange a date to performthis corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Acknowledge Receipt of this Safety Advisory Notice.
- Initiated
- 2020-07-14
- Root cause
- Device Design
- Terminated
- 2021-07-23
- Distribution
- Nationwide
- Recall number
- Z-2848-2020
- Recall status
- Terminated
- CDRH record ID
- 182683
- FDA classified
- 2020-08-22
- Recall event number
- 86116