Open original FDA record ↗Recall record
- FDA identifier
- Z-2852-2016
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Sep 21, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Software error. Incorrect values for the volume calculation from a freehand VOI at the customer site. In volume calculations of prostate as well as in volume calculations of liver were too high.
Source facts
- Action
- Siemens sent an Customer Letter dated August 5, 2016, to all affected customers. The letter identified the product the problem and the action need to be taken by the customer. The letter informed customers of a software change VB10B_HF02 that they have implemented to address several issues. If the device has been sold and therefore no longer in your possession, please forward this notice to the new owner. We would also request that you inform us of the identity of the device's new owner where possible. Please observe this safety notice and comply with the corresponding measure until the update has been fully completed. For further questions please call (610) 219-4834.
- Initiated
- 2016-08-05
- Root cause
- Software design
- Terminated
- 2017-08-22
- Distribution
- Distributed Nationwide
- Recall number
- Z-2852-2016
- Recall status
- Terminated
- CDRH record ID
- 149458
- FDA classified
- 2016-09-21
- Recall event number
- 75143