Open original FDA record ↗Recall record
- FDA identifier
- Z-2853-2011
- Company in FDA record
- Baxter Healthcare Corp.
- FDA classification date
- Jul 19, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
While replacing a defective pump head mechanism and performing calibration, Baxter's Service center may have failed to apply GLYPTOL (a red insulating varnish) on a potentiometer that helps maintain pump accuracy for detecting an air bolus.
Source facts
- Action
- Baxter Healthcare Corporation telephoned its customers on July 1, 2011 and sent a follow up "URGENT PRODUCT RECALL" letter dated July 6, 2011, informing them that the Glyptol (a red insulating varnish) may not have been applied on the Air-In-Line (AIL) circuit board. The letter described the product, problem and actions to be taken. The customers were instructed to locate the affected pumps, remove from service and return to Baxter for inspection as soon as possible. If GLYPTOL is found to be missing, recalibration will be performed and GLYPTOL will be applied per requirement before returning the pumps back to the customer at no cost. The customer was instructed to complete and sign the attached customer reply form. The letter states that if the COLLEAGUE Pumps were distributed to other facilities, please forward communication to the pump owner as appropriate. For any questions contact The Center for One Baxter at 1-800-422-9837 during the hours of 8:00 am to 5:00 pm Central Time, Monday through Friday.
- Initiated
- 2011-07-05
- Root cause
- Process control
- Terminated
- 2012-01-20
- Distribution
- Nationwide (USA) Distribution -- including the state of Pennsylvania.
- Recall number
- Z-2853-2011
- Recall status
- Terminated
- CDRH record ID
- 102087
- FDA classified
- 2011-07-19
- Recall event number
- 59270