Terumo Cardiovascular Systems received four (4) complaints from a single customer about tunnel breathing/inadequate insufflation with the VirtuoSaph Plus Endoscopic Vessel Harvesting System during cases in a limited production launch situation.
Teumo's investigation identified issues with the devices's capability to maintain pressure within the tunnel.
Source facts
Action
Terumo Cardiovascular Systems contacted the one customer to this device on June 10, 2010 by telephone.
There was one consignee. The firm contacted this one consignee by phone on 6/10/2010. There were 2 contacts at this one consignee. The recalling firm contacted both of the contacts on 6/10/2010. The consignee was asked to dispose of remaking inventory of devices. The consignee said had disposed of the product.
For information on this recall call (734) 741-6173
Initiated
2011-06-10
Root cause
Device Design
Terminated
2011-07-27
Distribution
Nationwide (USA) Distribution including the state of New York.
Recall number
Z-2854-2011
Recall status
Terminated
CDRH record ID
101568
FDA classified
2011-07-21
Recall event number
59189