Skip to content

Public FDA record

Z-2854-2026: Brand Name: WASPLab Product Name: WASPLab Model/Catalog Number: W087 Software Version: WebApp version 2.3

Recall record Z-2854-2026 for Copan WASP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2854-2026
Company in FDA record
Copan WASP
FDA classification date
Jul 31, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a software issue within the WebApp that potentially prevents the user to overwrite the results.

Source facts

Action
On 06/16/2026, an "URGENT MEDICAL DEVICE RECALL" Letter was provided to customers informing them that, A software bug has been identified in the WebApp software which, under specific conditions, may prevent the result suggested by PhenoMATRIX from being overwritten by the result selected by the user. As a consequence, the result transmitted by WASPLab to the LIS may be the result suggested by PhenoMATRIX, rather than the result selected by the user. Customers were informed and instructed: In date 2026/06/15 we released a WebApp software version including the bugfix of this issue. As a consequence, you are required to schedule with your distributor the software update to correct the issue and, in the meanwhile, to check the results sent to LIS in case you changed the result suggested by PhenoMATRIX. Complete the attached Customer Response Form and return to the Copan WASP Srl contact noted on the form so that Copan WASP Srl may acknowledge your receipt of this notification. For questions or further assistance email regulatory.wasp@copangroup.com
Initiated
2026-06-16
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the state of Arizona.
Recall number
Z-2854-2026
Recall status
Open, Classified
CDRH record ID
221230
FDA classified
2026-07-31
Recall event number
99302

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗