Skip to content

Public FDA record

Z-2855-2026: Nerivio Infinity. Model/Catalog Number: FGD000100. Software Version: FW Version 4.1.2.0, App version 7.2

Recall record Z-2855-2026 for THERANICA BIOELECTRONICS LTD. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2855-2026
Company in FDA record
THERANICA BIOELECTRONICS LTD
FDA classification date
Jul 31, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for the batter to have a thermal-runway event during charging, resulting in smoke and odor.

Source facts

Action
Firm issued a stop ship of product from distributors starting March 22, 2026 after learning of the reported event. These devices were released from hold on 4/30/2026, after the device firmware update was released on 4/19/26. Starting May 13, 2026, customers received in-app messaging which ensure that all affected users receive instructions regarding the required corrective action. Users were prompted to perform a software update affecting both the mobile application used to control the device and the device firmware. The firmware update permanently disables the device upon detection of an overheating event. The software update provides an enhanced warning mechanism if it detects a device malfunction.
Initiated
2026-03-23
Root cause
Component design/selection
Distribution
The product's entry point to the US is a 3PL warehouse from which it is being shipped to DME distributors or specialty pharmacy who perform the shipment to the end-user
Recall number
Z-2855-2026
Recall status
Open, Classified
CDRH record ID
221248
FDA classified
2026-07-31
Recall event number
99307

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.