Z-2856-2026: munevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart Glasses Inte

Recall record Z-2856-2026 for MUNEVO GMBH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2856-2026
Company in FDA record
MUNEVO GMBH
FDA classification date
Jul 31, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A software issue may cause the wheelchair's driving mechanism to fall out of calibration. When this occurs, the device may experience unexpected steering drift or imprecise control while driving the wheelchair. The wheelchair may veer from its intended path, which may cause in low-speed collision, tipping, or injuries to user or bystanders.

Source facts

Action
On June 19, 2026, the firm issued an "Urgent Medical Device Recall" letter to consignees, which detailed the hazard, mandatory boot-up safety protocols, and manual override steps. The firm has developed software patch Version 12.1.7 (v1), and 12.2.6 (R-net) (released patch on June 17 2026) which corrects the profile retention database loop, ensuring the device consistently locks onto the last saved user profile during boot. This software patch was released to all users on June 17, 2026. Customers should update the software as soon as possible. Until the update is installed, do not activate neutral calibration in the admin menu (only applicable to those with access to the admin menu.)
Initiated
2026-06-19
Root cause
Software Design Change
Distribution
Worldwide - US Nationwide distribution in the states of California, Florida, North Carolina, Texas, Utah, and Washington. The countries of United Kingdom, Spain, Italy, Germany, and Switzerland.
Recall number
Z-2856-2026
Recall status
Open, Classified
CDRH record ID
221303
FDA classified
2026-07-31
Recall event number
99328

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗