Z-2857-2026: Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System Model/Catalog Number: MC2, MC2.1 So

Recall record Z-2857-2026 for Oxos Medical Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2857-2026
Company in FDA record
Oxos Medical Inc
FDA classification date
Aug 3, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a defective circuit board that overheats

Source facts

Action
On 06/12/2026, an "URGENT: Medical Device Recall - Limited to Specific MC2 Units" letter was emailed to customers informing them that, due to faulty circuit boards limited units may experience localized overheating. As a result the firm released a software update (version 5.7.0) which will place a unit in a safe state if boards fail. Customers are instructed to: 1. Update units to software version 5.7.0 via over-air updates. Note, it is important to apply the software update to continue using the MC2 units 2. In addition to the software update, OXOS will replace the defective components with the cassette as part of a scheduled procedure. The MC2 must be returned to OXOS headquarters for replacement of affected components. Servicing will coordinate with customers to minimized any inconvenience. 3. Customer need to acknowledge receipt of the recall communication and complete the software update by calling OXOS Support Team at 866-749-4229 to schedule a time to perform the updated by emailing support#oxos.com, or accepting the update on their devices. For questions or assistance - contact Support Team at 866-749-4229 or email support@oxos.com.
Initiated
2026-06-11
Root cause
Process change control
Distribution
US Nationwide distribution in the states of FL, GA, IN, SC, TX, VA.
Recall number
Z-2857-2026
Recall status
Open, Classified
CDRH record ID
221166
FDA classified
2026-08-03
Recall event number
99263

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗