Open original FDA record ↗Recall record
- FDA identifier
- Z-2858-2026
- Company in FDA record
- Medical Action Industries, Inc. 306
- FDA classification date
- Aug 3, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Convenience kits contain needle free access device that has an improper seal which may result in doming or tearing of the silicone seal.
Source facts
- Action
- On May 21, 2026 Medical Action Industries issued a "Urgent: Medical Device" Recall notification via Email. Medical Action Industries asked consignees to take the following actions: 1. Identify, Segregate, and Quarantine all impacted kits listed in Table 1. 2. Add warning labels to affected AVID kits to ensure all users are aware of discarding the impacted component prior to use. 3. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. 4. Customers (end-users): If you have received product from a distributor, please contact the distributor or your sales representative with any questions you may have. 5. Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. 6. Complete the attached Recall Response Form and return by June 5, 2026.
- Initiated
- 2026-05-21
- Root cause
- Nonconforming Material/Component
- Distribution
- US Nationwide distribution in the state of CA.
- Recall number
- Z-2858-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221342
- FDA classified
- 2026-08-03
- Recall event number
- 99338