Skip to content

Public FDA record

Z-2858-2026: Brand Name: Medical Action Industries Product Name: Dialysis Change Tray Clear Sequence Model/Catalog Nu

Recall record Z-2858-2026 for Medical Action Industries, Inc. 306. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2858-2026
Company in FDA record
Medical Action Industries, Inc. 306
FDA classification date
Aug 3, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Convenience kits contain needle free access device that has an improper seal which may result in doming or tearing of the silicone seal.

Source facts

Action
On May 21, 2026 Medical Action Industries issued a "Urgent: Medical Device" Recall notification via Email. Medical Action Industries asked consignees to take the following actions: 1. Identify, Segregate, and Quarantine all impacted kits listed in Table 1. 2. Add warning labels to affected AVID kits to ensure all users are aware of discarding the impacted component prior to use. 3. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. 4. Customers (end-users): If you have received product from a distributor, please contact the distributor or your sales representative with any questions you may have. 5. Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. 6. Complete the attached Recall Response Form and return by June 5, 2026.
Initiated
2026-05-21
Root cause
Nonconforming Material/Component
Distribution
US Nationwide distribution in the state of CA.
Recall number
Z-2858-2026
Recall status
Open, Classified
CDRH record ID
221342
FDA classified
2026-08-03
Recall event number
99338

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗