Open original FDA record ↗Recall record
- FDA identifier
- Z-2859-2026
- Company in FDA record
- Innovasis, Inc
- FDA classification date
- Aug 3, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Pedicle screw system intended to provide immobilization/stabilization of spinal segments as adjunct to thoracic/lumbar/sacral spine fusion may have anti-torque instrument with undersized condition which could prevent mating tip from engaging with the pedicle screw tulip and rod during tightening, which may lead to surgical delay, intervention/alternative methods to perform final tightening step.
Source facts
- Action
- On 5/22/2026, recall notices were emailed to distributors who were informed of the following: 1) Firm personnel will coordinate product replacement with customers. 2) Once all affected parts have been returned, a Product Removal Attestation Form will be sent to customers. This form must be filled in, signed and returned for regulatory purposes. If you have any questions, contact firm Customer Support at support@innovasis.com 801-261-2242 Mon-Fri 8:00am - 5:00pm MST.
- Initiated
- 2026-05-22
- Root cause
- Process design
- Distribution
- US Nationwide distribution in the states of TX, CA.
- Recall number
- Z-2859-2026
- Recall status
- Open, Classified
- CDRH record ID
- 220999
- FDA classified
- 2026-08-03
- Recall event number
- 99188