Skip to content

Public FDA record

Z-2859-2026: Anti-Torque, Vector, Part: LS-749 part of Vector Pedicle Screw System

Recall record Z-2859-2026 for Innovasis, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2859-2026
Company in FDA record
Innovasis, Inc
FDA classification date
Aug 3, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Pedicle screw system intended to provide immobilization/stabilization of spinal segments as adjunct to thoracic/lumbar/sacral spine fusion may have anti-torque instrument with undersized condition which could prevent mating tip from engaging with the pedicle screw tulip and rod during tightening, which may lead to surgical delay, intervention/alternative methods to perform final tightening step.

Source facts

Action
On 5/22/2026, recall notices were emailed to distributors who were informed of the following: 1) Firm personnel will coordinate product replacement with customers. 2) Once all affected parts have been returned, a Product Removal Attestation Form will be sent to customers. This form must be filled in, signed and returned for regulatory purposes. If you have any questions, contact firm Customer Support at support@innovasis.com 801-261-2242 Mon-Fri 8:00am - 5:00pm MST.
Initiated
2026-05-22
Root cause
Process design
Distribution
US Nationwide distribution in the states of TX, CA.
Recall number
Z-2859-2026
Recall status
Open, Classified
CDRH record ID
220999
FDA classified
2026-08-03
Recall event number
99188

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗