Z-2863-2011: Trauma Stryker Medical Stretchers Model 1037 Intended to transport patients in a horizontal position.

Recall record Z-2863-2011 for Stryker Medical Division of Stryker Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2863-2011
Company in FDA record
Stryker Medical Division of Stryker Corporation
FDA classification date
Jul 22, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The firm has determined that the brake rod assemblies of various stretchers that were built with a 5th wheel were manufactured with roll pins that did not meet the materials specifications. Some roll pins may have fractured during the assembly process and could fall out of the brake rod assemblies, rendering the brake or steer systems inoperable from certain pedals or causing a reduction in overa

Source facts

Action
Stryker Medical sent an "Urgent Medical Device Correction" dated July 19 ,2011 to all affected customers. The letter included; the name and model numbers of the stretchers, the reason for the correction and dates of manufacture. It contained the instructions to check the braking system and if found to be inoperable, to remove it from service, and if operable to check prior to each use. Also, information regarding contacting the recipients of any loaned or sold stretchers to relay the information in the letter and Medwatch reporting information. A self addressed postcard to confirm receipt of the letter was also included. For additional information call (269) 389-6604.
Initiated
2011-07-19
Root cause
Nonconforming Material/Component
Terminated
2013-08-29
Distribution
Worldwide Distribution: Nationwide (USA) including the states of NY, MI, LA and CA ,and the countries of Canada, China, Spain, Japan, Korea, United Kingdom, South Africa.
Recall number
Z-2863-2011
Recall status
Terminated
CDRH record ID
102120
FDA classified
2011-07-22
Recall event number
59201

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗