Specific lots of ALTERA implants have internal components that may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy.
Source facts
Action
A customer letter dated July 22, 2020 was mailed to customers. In this letter customers are instructed to immediately examine their inventory and determine if they
have any of the affected implants. Alternatively, they may return all ALTERA inventory and Globus will sort out the affected implants and replace inventory with conforming implants. Report any adverse effects believed to be associated with the product to Globus Medical immediately by filling out the attached Response Card.
Initiated
2020-07-22
Root cause
Nonconforming Material/Component
Terminated
2021-08-19
Distribution
Worldwide Distribution
Recall number
Z-2868-2020
Recall status
Terminated
CDRH record ID
182920
FDA classified
2020-08-22
Recall event number
86164