GEHC Surgery became aware that some users of the 9800 and 9900 fluoroscopic mobile C-Arms (referred to as the system here on) within the United States had requested that a Field Service Engineer (FSE) change the dose value on the live update screen from AKR / CAK to DAP. All systems sold to customers in the US that leave GEHC Surgery (the place of manufacture) are configured to display AKR / CAK when they leave the factory. An FSE changing the dose display, per the customer request, to display DAP instead does not comply to 21CFR 1020.32(k), which requires fluoroscopic equipment manufactured on or after June 10, 2006 to display continuous updates of AKR and CAK values.
Source facts
Initiated
2020-05-08
Root cause
Radiation Control for Health and Safety Act
Distribution
Nationwide
Recall number
Z-2872-2020
Recall status
Open, Classified
CDRH record ID
183013
FDA classified
2020-08-27
Recall event number
86201