- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
On May 17, 2010, Biomet Microfixation issued a Safety Alert on their Thinflap System Cross Drive, Self Drilling Screw, 1.5x4mm, Model #: 91-6704, Lot#: 173770. The product is labeled incorrectly. Some packages of 91-6704 lot 173770 are labeled as cross drive screws, but contain center drive screws. All affected customers were notified of the recall on May 17, 2010.
Source facts
- Action
- Biomet Microfixation notified distributors and customers by letter, email and telephone calls. Distributors and customers were instructed to fill out and return the Inventory Reconciliation sheet, via Fax to 904-741-9425. Distributors were also instructed to notify their customers if the product was distributed further. For questions regarding this recall customers were instructed to call 1-800-874-7711 or 904-741-4400, ext. 468.
- Initiated
- 2010-05-17
- Root cause
- Labeling mix-ups
- Terminated
- 2011-10-12
- Distribution
- Worldwide Distribution - USA including AL, CA, GA, HI, IL, MO, NC, NY, TX and the country of Japan.
- Recall number
- Z-2873-2011
- Recall status
- Terminated
- CDRH record ID
- 100808
- FDA classified
- 2011-07-22
- Recall event number
- 59002