Z-2873-2018: GE Healthcare Optima CT660. The systems are intended for head, whole body, cardiac and vascular X-ray Co

Recall record Z-2873-2018 for GE Healthcare Japan Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2873-2018
Company in FDA record
GE Healthcare Japan Corporation
FDA record creation date
Aug 17, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

GE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to Service Engineers working inside the gantry.

Source facts

Action
The firm, GE Healthcare, sent an " URGENT MEDICAL DEVICE CORRECTION" letter dated 4/27/2018 to its Customers on about 04/27/2018. The letter describes the product, problem and actions to be taken. The customers were Instructed to: download and provide the updated Service Manual to the organization that services the equipment on site and informed customers that GE Healthcare would be correcting all affected devices at no cost. Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or the local Sales Representative. GE Healthcare expanded the recall to include an additional product and model number. Customers were notified on about 08/09/2018 via URGENT MEDICAL DEVICE CORRECTION letter. Instructions were the same as the original notification.
Initiated
2018-04-27
Root cause
Employee error
Terminated
2020-08-21
Distribution
Worldwide Distribution: US (Nationwide) distribution to states of: AK, AL, CA, CO, CT, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, RI, SC, SD, TN, TX, UT, WI, WV, WY; and countries of:: ALGERIA, ARGENTINA, ARMENIA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BOLIVIA, BOSNIA AND HERZEGOVINA, BRAZIL, BULGARIA, CAMEROON, Canada, CHILE, CHINA, COLOMBIA, Cote D'Ivoire, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, ETHIOPIA, FINLAND, FRANCE, French Guiana, GERMANY, GHANA, GREECE, GUADELOUPE, GUATEMALA, HONDURAS, Hong Kong, HUNGARY, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, Korea (Republic of), KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, MALAYSIA, Mali, MALTA, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NEW , ZEALAND, NIGERIA, NORWAY, OMAN, PAKISTAN, Palestine, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, PUERTO RICO, REUNION, ROMANIA, Russia, RWANDA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TANZANIA, THAILAND, TUNISIA, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, UZBEKISTAN, VENEZUELA, VIETNAM, and YEMEN.
Recall number
Z-2873-2018
Recall status
Terminated
CDRH record ID
166439
FDA record created
2018-08-17
Recall event number
80563

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗