Open original FDA record ↗Recall record
- FDA identifier
- Z-2874-2026
- Company in FDA record
- Merge Healthcare, Inc.
- FDA classification date
- Aug 4, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Thumbnails from a previously opened patient study may be displayed in the Thumbnail Bar while a different patient study is currently open.
Source facts
- Action
- Merge Healthcare issued an Urgent Field Safety Notice (Revision 2) to its consignees on 6/30/2026 via email. The notice explained the issue and when it may occur, risk to patient, and requested an immediate workaround, a configuration change, be performed to disable automatic refresh of the Thumbnail Bar until a software update is available. The notice also directions the consignee to forward the notice to any individuals within their organization that may need to be aware of this notification. Customers should alert any other affiliated parties that may be affected by this Urgent Field Safety Notice. The notice was followed by telephone calls beginning on 7/6/2026 to ensure the notice was received. For more information, please contact Merge Healthcare Support at 1-877-741-5369 (available 24 hours a day, 7 days a week) or visit our website: https://www.merative.com/support/merge.
- Initiated
- 2026-06-30
- Root cause
- Software Design Change
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, and UK.
- Recall number
- Z-2874-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221464
- FDA classified
- 2026-08-04
- Recall event number
- 99391