Open original FDA record ↗Recall record
- FDA identifier
- Z-2877-2018
- Company in FDA record
- TriMed Inc.
- FDA record creation date
- Aug 18, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Two instrument holes located on the side of the plate, which act to align the saw guide are located on the wrong side of the plate
Source facts
- Action
- Tri-Med issued notification on April 16, 2018 requesting: Determine the lot number of all Ulnar Osteotomy Compression Plates, Short (UOCP-S) .If you have UOCP-S plates with the lot numbers listed (10003, 10017): o Remove the UOCP-S plates from the tray and quarantine it. Do not use this plate. Contact TriMed at quality@trimedortho.com or call 800-633-7221. TriMed will arrange have the plate(s) returned to the company and replaced. Return the Recall Response card attached to this letter to TriMed as soon as possible. Contact TriMed via telephone or email as per below if you have any questions: Telephone*: 800-633-7221, Email: quality@trimedortho.com. TriMed will replace the potentially defective plates with plates that have been manufactured and confirmed to meet TriMeds specifications.
- Initiated
- 2018-04-16
- Root cause
- Process control
- Terminated
- 2021-01-28
- Distribution
- Worldwide Distribution in the state of TX and countries of Australia, China Sweden
- Recall number
- Z-2877-2018
- Recall status
- Terminated
- CDRH record ID
- 165588
- FDA record created
- 2018-08-18
- Recall event number
- 80487