When attempting to recapture the 32-mm device, the tip of the delivery sheath may fold or buckle, resulting in increased retraction force and difficulty or failure to recapture the device.
Source facts
Action
Coherex Medical notified customers via "URGENT FIELD SAFETY NOTICE
Coherex WaveCrest Left Atrial Appendage (LAA) Occlusion System (32 mm)
Catalog No: WCR1523, Lot Number: All" letter dated May 7, 2018.
The letter provided an overview, and important information about the product description and codes, lots subject to removal, why customers were being contacted, What actions are required/instructions, available assistance/contact information, and Reply Form.
Initiated
2018-05-07
Root cause
Component design/selection
Terminated
2020-11-14
Distribution
No U.S. Distribution. International Distribution only.
Recall number
Z-2878-2018
Recall status
Terminated
CDRH record ID
166222
FDA record created
2018-08-30
Recall event number
80720