Skip to content

Public FDA record

Z-2878-2026: GE Healthcare Portrait Core Services SW V 1.2.0, REF 5700592. patient monitoring system

Recall record Z-2878-2026 for GE Medical Systems Information Technologies Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2878-2026
Company in FDA record
GE Medical Systems Information Technologies Inc
FDA classification date
Aug 4, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A software vulnerability may allow unauthorized persons to gain privileged access to the device, which could result in loss of patient monitoring, missed patient alarms, and/or access to Protected Health Information (PHI).

Source facts

Action
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION to its consignees on 7/10/2026 via traceable means. The notice explained the safety issue, and requested the following: "Actions to be taken by Customer/User You can continue to use Portrait Mobile Monitoring Solution with Portrait Core Services SW V1.2.0, pending GE HealthCare correction. If you need help to identify if your Portrait Mobile Monitoring Solution has Portrait Core Services SW 1.2.0, contact GE HealthCare Service. Ensure that your facility follows all hospital networking and cybersecurity policies and procedures along with the security considerations outlined in the GE HealthCare Portrait Mobile Monitoring Solution Privacy and Security Manual p/n 5970026. Please retain this document for your records." For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or a local Service Representative.
Initiated
2026-07-10
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution in the state of MI, and the countries of Czech Republic, Finland, Germany, Netherland, Qatar, UAE.
Recall number
Z-2878-2026
Recall status
Open, Classified
CDRH record ID
221631
FDA classified
2026-08-04
Recall event number
99475

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗