Z-2880-2020: CARTO 3 System (Software Version V7.1.80), CARTO VISITAG Module - Product Usage: FGThe intended use of th

Recall record Z-2880-2020 for Biosense Webster (israel), Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2880-2020
Company in FDA record
Biosense Webster (israel), Ltd.
FDA record creation date
Aug 24, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software defect may result in disapperance of tag sites during recalculation and lead to additional ablation sites. This may lead to prolongation of the procedure and, in extremely rare circumstances, cardiac perforation.

Source facts

Action
On 06/22/2020, the firm notified affected US customers via mail of a hospital letter, "URGENT Medical Device Correction", which indicates the following: This product is not being removed from the market and does not need to be returned. A software service pack will be provided by the firm. Your representative will notify you when it is available and will arrange for installation of the service pack. In the meantime, the firm has identified certain mitigations preventing the issue from occurring: If your system is equipped with the VISITAG SURPOINT Module License: - Always enable the VISITAG SURPOINT Module and ensure that the External Processing Unit (EPU) is connected before starting a case. - When using review mode, ensure that the EPU is connected before entering the study and verify that a green connection indication is present on the screen. - When CARTO VISITAG recalculation was performed without the VISITAG SURPOINT Module enabled, recalculate the CARTO VISITAG results with the VISITAG SURPOINT Module by physically reconnecting the EPU, then exit and continue the study and verify that a green connection indication is present on the screen. If your system is not equipped with a VISITAG SURPOINT Module License: - Avoid triggering CARTO VISITAG recalculations as much as possible. - During the procedure consider applying a manual tag for each ablation location by acquiring an ablation point in addition to the CARTO VISITAG Module marking the ablation locations. - If a CARTO VISITAG tag disappears, the CARTO 3 system will still provide an indication of ablation at this location for the manually acquired points. - During offline review of a previous study, be aware that some CARTO VISITAG tags may have disappeared. For next steps, 1. Please review this letter carefully and share it with anyone in your facility that needs to be informed. 2. Please complete, sign, and return the Business Reply Form to the fim via email. The firm notif
Initiated
2020-06-22
Root cause
Software design
Terminated
2022-01-06
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AK, AZ, AR, CA, CO, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WV, and WI *Three (3) U.S. government consignees. The countries of Australia, Austria, Belgium, Chile, China, Colombia, Czechia, Denmark, Finland, France, Germany, Hong Kong, Hungary, Ireland, Israel, Italy, Japan, Latvia, Netherlands, Norway, Poland, Portugal, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, United Kingdom.
Recall number
Z-2880-2020
Recall status
Terminated
CDRH record ID
182136
FDA record created
2020-08-24
Recall event number
85910

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗